可穿戴式光纤光栅步态监测装置的分析方法

闫希研,  冯艳,  王昊祥,  张华

闫希研, 冯艳, 王昊祥, 张华. 可穿戴式光纤光栅步态监测装置的分析方法[J]. 上海工程技术大学学报, 2026, 40(3): 348-354. doi: 10.12299/jsues.25-0075
引用本文: 闫希研, 冯艳, 王昊祥, 张华. 可穿戴式光纤光栅步态监测装置的分析方法[J]. 上海工程技术大学学报, 2026, 40(3): 348-354. doi: 10.12299/jsues.25-0075
YAN Xiyan, FENG Yan, WANG Haoxiang, ZHANG Hua. Analysis method of wearable fiber grating gait monitoring device[J]. Journal of Shanghai University of Engineering Science, 2026, 40(3): 348-354. doi: 10.12299/jsues.25-0075
Citation: YAN Xiyan, FENG Yan, WANG Haoxiang, ZHANG Hua. Analysis method of wearable fiber grating gait monitoring device[J]. Journal of Shanghai University of Engineering Science, 2026, 40(3): 348-354. doi: 10.12299/jsues.25-0075

可穿戴式光纤光栅步态监测装置的分析方法

doi: 10.12299/jsues.25-0075
基金项目: 上海市地方院校能力建设项目(61763030)
详细信息
    作者简介:

    闫希研(1999 − ),男,硕士生,研究方向为光纤智能感知的研究。E-mail:15029492776@163.com

    通讯作者:

    冯 艳(1976 − ),女,教授,博士,研究方向为光纤光栅传感、智能机器人触觉感知等。E-mail:xmfy0833@sina.com

  • 中图分类号: TN253;TP212

Analysis method of wearable fiber grating gait monitoring device

  • 摘要: 根据光纤布拉格光栅(fiber Bragg grating, FBG)传感器的特性建立FBG传感器和LabVIEW步态监测系统的协同精确监测方法。按照膝盖和小腿的不同生理结构设计穿戴式FBG膝腿运动监测装置。骨骼模型实验验证差动方法消除温度误差的合理性;通过人体膝盖和小腿的标定实验分别得到膝盖的平均灵敏度为3.85 pm/(°),小腿的平均灵敏度4.05 pm/(°),充分证明FBG穿戴装置和LabVIEW步态监测系统可以监测出膝盖和小腿的运动角度。LabVIEW步态监测系统测得的理论步距与实际步距的相对误差为5.525%,理论步高与实际步高的相对误差为4.750%。实验研究表明,LabVIEW步态监测系统和FBG穿戴装置可为监测人体步态参数提供技术支持。
  • 图  1  光纤光栅传感器的工作原理图

    Figure  1.  Working principal diagram of fiber grating sensor

    图  2  监测示意图

    Figure  2.  Monitoring schematic diagram

    图  3  可穿戴装置的制备流程

    Figure  3.  Preparation process of wearable device

    图  4  LabVIEW 步态监测系统

    Figure  4.  LabVIEW gait monitoring system

    图  5  骨骼模型图

    Figure  5.  Skeleton model diagram

    图  6  膝盖实验

    Figure  6.  Knee experiment

    图  7  小腿实验

    Figure  7.  Lower leg experiment

    图  8  步高−步距分析

    Figure  8.  Step height-step analysis

    表  1  运动波长差动式运算数据

    Table  1.   Differential operation data of moving wavelength

    监测温度 波长/nm 相对误差/% 平均误差/%
    20 ℃ 25 ℃ 30 ℃ 20~25 ℃ 25~30 ℃ 30~20 ℃
    0° 0.879 5 0.878 2 0.883 2 0.148 0.566 0.419 0.378
    5° 0.892 6 0.895 0 0.901 2 0.268 0.688 0.954 0.637
    10° 0.905 6 0.906 3 0.910 3 0.077 0.439 0.516 0.344
    15° 0.922 4 0.919 5 0.930 5 0.314 1.182 0.871 0.789
    20° 0.945 6 0.950 1 0.949 2 0.474 0.095 0.379 0.316
    下载: 导出CSV

    表  2  步高上升阶段的步态理论数据和实际数据

    Table  2.   Theoretical and practical data of gait during step-height ascent

    测量
    次数
    $ {\theta }_{1} $/(°) $ {\theta }_{2} $/(°) 理论
    步距
    /cm
    实际
    步距
    /cm
    理论
    步高
    /cm
    实际
    步高
    /cm
    步距
    误差
    /%
    步高
    误差
    /%
    1 0 0 0 0 0 0 0 0
    2 5 3.4 11.7 12.1 0.23 0.25 3.3 8.0
    3 10 7.1 23.7 25.3 0.92 0.99 6.3 7.1
    4 15 11.4 36.4 38.9 2.16 2.20 6.4 1.8
    5 20 15.1 48.0 51.1 3.81 3.89 6.1 2.1
    下载: 导出CSV
  • [1] Song Y, Liu T J, Wang F L, et al. Wearable capacitive pressure sensors based on porous sponge dielectric materials for gait analysis and identity recognition[J] . Advanced Materials Technologies, 2025, 10(11): 2401768. doi: 10.1002/admt.202401768
    [2] Cheng Z H, Wen Y Z, Xie Z N, et al. A multi-sensor coupled supramolecular elastomer empowers intelligent monitoring of human gait and arch health[J] . Chemical Engineering Journal, 2025, 504: 158760. doi: 10.1016/j.cej.2024.158760
    [3] Lallès A P, Moucheboeuf G, Doat E, et al. IMU calibration effect on lower limbs kinematics against optical motion capture in post-stroke gait[J] . IRBM, 2025, 46(1): 100873. doi: 10.1016/j.irbm.2024.100873
    [4] Zhao L L, Jia S C, Fang C Y, et al. Machine learning-assisted wearable triboelectric-electromagnetic sensor for monitoring human motion feature[J] . Chemical Engineering Journal, 2025, 503: 158637. doi: 10.1016/j.cej.2024.158637
    [5] Gao Y Y, Zong R S, Feng J Y, et al. Flexible self-powered and self-sensing shoes based on aeroelastic structure for application in human motion monitoring[J] . Journal of Power Sources, 2025, 638: 236626. doi: 10.1016/j.jpowsour.2025.236626
    [6] 胡高童. 基于无线智能鞋垫的人体运动状态识别方法研究[D] . 武汉: 武汉纺织大学, 2024.
    [7] 谢俊枭, 赵欢, 曹军义, 等. 帕金森病步态的可穿戴多源数据采集与特征识别方法[J] . 西安交通大学学报, 2023, 57(6): 29 − 38.
    [8] 彭爽. 基于惯性传感器的脑卒中患者家用步态采集系统研究[D] . 重庆: 重庆医科大学, 2020.
    [9] Gui X, Zeng F H, Gao J, et al. Detection of Aircraft wing icing and de-icing by optical fiber sensing with FBG array[J] . Measurement, 2025, 247: 116748. doi: 10.1016/j.measurement.2025.116748
    [10] 秦驰. 用于足踝矫形器的埋入式光纤传感器[D] . 哈尔滨: 哈尔滨工程大学, 2024.
    [11] Song X B, Fan Y W, Tang X Y. FBG-based wearable sensors and devices in the healthcare field: A review[J] . Optics & Laser Technology, 2025, 181: 111920. doi: 10.1016/j.optlastec.2024.111920
    [12] Zhou X, Wang W J, Hui Y P, et al. Real-time human posture recognition using embedded optical fiber sensors and machine learning[J] . IEEE Transactions on Instrumentation and Measurement, 2025, 74: 2508609.
    [13] Yi P, Luo B B, Wu D C, et al. Design and gesture recognition of wrist joint posture sensor based on fiber Bragg gratings[J] . IEEE Sensors Journal, 2024, 24(23): 39050 − 39058. doi: 10.1109/JSEN.2024.3481013
    [14] Zhang C, Zhou H Y, Nong H C, et al. FBG-SenseGlove: a digital twin data glove for underwater gesture recognition using deep learning[J] . IEEE Sensors Journal, 2025, 25(2): 8347 − 8354. doi: 10.1109/jsen.2025.3532601/mm1
    [15] Qiu Z C, Zhang Y F, Fan X Y, et al. Ultra-low frequency three-axis FBG accelerometer for seismic monitoring based on multiple flexible beams[J] . Measurement, 2025, 248: 116947. doi: 10.1016/j.measurement.2025.116947
    [16] 王昊祥, 冯艳, 潘睿智, 等. 光纤光栅波电转换的人机仿生跟随[J] . 光学精密工程, 2023, 31(23): 3414 − 3425. doi: 10.37188/OPE.20233123.3414